The IEC61000-4-3- standard defines that the harmonics generated by the amplifier shall be such that any measured field strength in the UFA at each harmonic frequency shall be at least 6 dB below that of the fundamental frequency.
The RadiField is an integrated system which includes amplifiers and antennae of which the harmonics have been measured. Test reports show that the actual harmonics are at least 15 dB (typical) lower at all frequencies. At harmonic frequencies above 15 GHz the suppression is even more than 30 dB.
Conclusion: The RadiField by far meets the harmonic specification of – 6dBc.