Raditeq | FAQ
FAQ Categories
Find you Answer!
Welcome to the Raditeq FAQ page! Whether you’re a new customer or a seasoned user, this section is designed to help you find answers to common questions about our products, services, and technologies. Troubleshooting has never been easier as we’ve compiled the most frequently asked questions to ensure you have all the support you need. If you don’t find the answer you’re looking for here, our dedicated support team is always ready to assist. Explore below and get the insights you need to make the most out of Raditeq’s cutting-edge solutions!
Search your question
RadiCentre | Modular Test Systems
CTR1004 | CTR1009 | When all mains cables are connected and the mains switches are set to “ON”, simple touch the front screen of the RadiCentre®.
CTR2008A | When using a RadiCentre® CTR2008A press the “on” button on the front of the RadiCentre®
CTR1001S | When using a RadiCentre CTR1001S (single slot). Simply turn on the mains


RadiField | Electric Field Generators
The RadiField is operated through the power supply and communication plug-in card (model PSU2400A). This plug-in card is installed in the RadiCentre (Pro) modular test system and occupies two slots. The PSU2400A has a mains power supply connector, mains ON/OFF switch and one SMA input for applying the RF carrier signal. The N-type output connector has four (4) different functions. First, it transfers the RF carrier signal to the RadiField Triple A unit. Next to this, it provides the 55VDC voltage supply and arranges bi-directional communication to the RadiField Triple A unit. Finally, it can automatically control the horizontal / vertical polarization of the RadiField Triple A unit when it is used in combination with the RadiTower model RTW2000A with integrated polarizer.


When installing the RadiField for the first time or after any disconnect it might be that the internal polarizer motor of the RadiTower has “forgotten” its settings. Therefore the current polarization need to be redefined. This is done through the RadiCentre screen.
Select under the RadiField Tab on the RadiCentre the polarization setup. Store the current position as Horizontal or Vertical (which ever one is applicable). Then move the mast to the alternate state (so from horizontal to vertical or visa versa). and Store the second position.
Step-by-step guide how to setup the polarizer
1. How do I start the setup for the RadiField polarizer?
Navigate to the Setup option on the RadiField display located within the RadiCentre.

2. How do I set the RadiField to the horizontal position?
Ensure the polarizer’s linear actuator is extended:
- Repeatedly press “Extend (CCW) Course” and “Extend (CCW) Fine” until the RadiField is in the horizontal position.
- Then, press “Store: Horizontal position” to save this position.
3. How can I confirm the RadiField is in the horizontal position?
After pressing “Store: Horizontal position,” the status should update to show that the RadiField is in the horizontal position.

4. How do I set the RadiField to the vertical position?
Press the “Retract (CW)” button repeatedly, turning the RadiField until it reaches a vertical position.
- Once in the vertical position, press “Store: Vertical position” to save it.
5. How can I confirm the RadiField is in the vertical position?
After pressing “Store: Vertical position,” the status should update to indicate the RadiField is in the vertical position.
6. How do I test the polarization switch?
Return to the main RadiField screen. The polarization should indicate “Vertical.” When you press “Horizontal,” the RadiField should rotate to the horizontal position.



RadiSense | Electric Field Probe
When starting up the the RadiSense® and measuring at low E-field levels or frequency the reading can be unstable. This is due to the internal filtering in the field probe. The filtering is set as a standard to the “Dynamic” setting. This means that it is searching for the most stable reading. In cases of low test levels and/or frequency it is impossible to find this stable reading for the probe and the filter setting needs to be changed from Dynamic to Filter 6 or higher. This will make the reading stable for read-out.
This can be done through RadiMation under the advanced settings of the probe

The filter setting can also be changed using the RadiCentre® CTR1004 and CTR1009 using the touch screen.



General Questions
Request general information
If you have general questions about all products or our service surrounding them please leave a question here or mail us!