ISO17025 accredited calibration for RadiSense® RSS3018U
ISO17025 accredited calibration performed at Kiwa Dare Calibrations in accordance to the international standard IEC 61000-4-3 (2006) 3e ed. + A1 (2007) + A2 (2010).
Measured frequency points:
Min/Max level: > 3 V/m and < 200 V/m till 150 MHz
Min/Max level: > 3 V/m and < 100 V/m from 150 MHz
10 kHz – 100 kHz: 10 kHz steps
100 kHz – 1 MHz: 100 kHz steps
1 MHz – 10 MHz: 1 MHz steps
10 MHz – 100 MHz: 10 MHz steps
100 MHz – 1 GHz: 50 MHz steps
1 GHz – 6 GHz: 200 MHz steps
6 GHz – 18 GHz: 500 MHz steps
Calibrated under RvA accreditation:
– Frequency response three probe orientations @ 20V/m
– Linearity for three probe orientations @ 100 MHz from – 6dB to + 6dB of the reference field level (1 dB steps)
– Anisotropy three probe orientations @ 100kHz; 1MHz; 10 MHz; 100MHz; 1GHz and 10GHz.
The calibration certificate is provided in PDF format and all calibration correction data is delivered in Excel format.
Please contact if we do not cover your industry.