ISO17025 accredited calibration for RadiSense® 2010 series
ISO17025 accredited calibration performed at Kiwa Dare Calibrations in accordance to the international standard IEC 61000-4-3 (2006) 3e ed. + A1 (2007) + A2 (2010).
Measured frequency points:
Min/Max level: > 3 V/m and < 200 V/m till 150 MHz
Min/Max level: > 3 V/m and < 100 V/m from 150 MHz
10 kHz – 100 kHz: 10 kHz steps
100 kHz – 1 MHz: 100 kHz steps
1 MHz – 10 MHz: 1 MHz steps
10 MHz – 100 MHz: 10 MHz steps
100 MHz – 1 GHz: 50 MHz steps
1 GHz – 6 GHz: 200 MHz steps
6 GHz – 12.5 GHz: 500 MHz steps
Calibrated under RvA accreditation:
– Frequency response three probe orientations @ 20V/m
– Linearity for three probe orientations @ 100 MHz from – 6dB to + 6dB of the reference field level (1 dB steps)
– Anisotropy three probe orientations @ 100kHz; 1MHz; 10 MHz; 100MHz; 1GHz and 10GHz.
The calibration certificate is provided in PDF format and all calibration correction data is delivered in Excel format.
Please contact if we do not cover your industry.
E2000 Connector Saver Assembly
Factory calibration, field probe 10 GHz
Factory calibration, field probe 18 GHz