ISO17025 accredited (RvA) calibration certificate for RadiSense® 26/40 GHz

    Direct supportCall +31 348 - 200 100Add to Quote

    ISO17025 accredited calibration for RadiSense®  series 26 GHz & 40 GHz

    ISO17025 accredited calibration performed at DARE!! Calibrations in accordance to the international standard IEC 61000-4-3 (2006) 3e ed. + A1 (2007) + A2 (2010).

     

    Specifications
    • The polarisation of the E-field is vertical.
    • The E-field is measured isotropic.

     

    Measured frequency points:
    Min/Max level: > 3 V/m and < 200 V/m till 150 MHz
    Min/Max level: > 3 V/m and < 100 V/m from 150 MHz
    10 kHz – 100 kHz: 10 kHz steps
    100 kHz – 1 MHz: 100 kHz steps
    1 MHz – 10 MHz: 1 MHz steps
    10 MHz – 100 MHz: 10 MHz steps
    100 MHz –  1 GHz: 50 MHz steps
    1 GHz – 6 GHz: 200 MHz steps
    6 GHz – 12.5 GHz: 500 MHz steps

    6 GHz – 18 GHz: 500 MHz steps

    26,5 GHz

    40 GHz

    Calibrated under RvA accreditation:
    – Frequency response three probe orientations @ 20V/m
    – Linearity for three probe orientations @ 100 MHz from – 6dB to + 6dB of the reference field level (1 dB steps)
    – Anisotropy three probe orientations @ 100kHz; 1MHz; 10 MHz; 100MHz; 1GHz and 10GHz. 

    The calibration certificate is provided in PDF format and all calibration correction data is delivered in Excel format.

    Industries